CVE-2014-5434 Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter a...

CriticalCVSS 9.8 EPSS 1.6%
Published
Updated
Exploit status
No public exploit known

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

Upgrade path

2 packages: 0 fixed.

PackageAffectedFixed inStatus
baxter SIGMA Spectrum Infusion System = 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 No fix listed Versions listed
baxter sigma spectrum infusion system firmware = 6.05 No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
baxter sigma spectrum infusion system firmware = 6.05 No fix listed
baxter SIGMA Spectrum Infusion System = 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 No fix listed

References