CVE-2015-1012 Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use,....
HighCVSS 7.5
EPSS 0.77%
- Published
- Updated
- Exploit status
- No public exploit known
Description
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| pfizer lifecare pca infusion system firmware | <= 5.0 | No fix listed |
| hospira lifecare pca infusion system | = <= 5.0 | No fix listed |
References
- ics-cert.us-cert.gov mitigation third-party-advisory us-government-resource