CVE-2015-3459 The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration...

HighCVSS 10.0 EPSS 5.1%
Published
Updated
Exploit status
No public exploit known

Description

The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.

Affected products

ProductAffected versionsFixed in
hospira lifecare pcainfusion firmware <= 5.0 No fix listed
hospira lifecare pca3 All versions No fix listed
hospira lifecare pca5 All versions No fix listed