CVE-2015-3956 Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pum...
CriticalCVSS 9.8
EPSS 0.95%
- Published
- Updated
- Exploit status
- No public exploit known
Description
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| pifzer plum a+ infusion system firmware | <= 13.4 | No fix listed |
| pifzer plum a+3 infusion system firmware | <= 13.6 | No fix listed |
| pifzer symbiq infusion system firmware | <= 3.13 | No fix listed |
| hospira Plum A+ Infusion System | = <= 13.4 | No fix listed |
| hospira Plum A+3 Infusion System | = <= 13.6 | No fix listed |
| hospira Symbiq Infusion System | = <= 3.13 | No fix listed |
References
- ics-cert.us-cert.gov mitigation third-party-advisory us-government-resource