CVE-2017-12716 Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionall...

MediumCVSS 6.5 EPSS 0.24%
Published
Updated
Exploit status
No public exploit known

Description

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

Affected products

ProductAffected versionsFixed in
abbott accent firmware < f0b.0e.7e f0b.0e.7e
abbott anthem firmware < f0b.0e.7e f0b.0e.7e
abbott accent mri firmware < f10.08.6c f10.08.6c
abbott accent st firmware < f10.08.6c f10.08.6c
Abbott Laboratories Accent and Anthem = All versions of pacemakers manufactured prior to August 28, 2017 No fix listed