CVE-2017-12724 A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentia...
HighCVSS 8.1
EPSS 1.3%
- Published
- Updated
- Exploit status
- No public exploit known
Description
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.
Upgrade path
1 package: 0 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.1; = 1.5; = 1.6 | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.1 | No fix listed |
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.5 | No fix listed |
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.6 | No fix listed |
References
- 100665 third-party-advisory vdb-entry
- ics-cert.us-cert.gov third-party-advisory us-government-resource