Skip to content

CVE-2017-12724 A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentia...

HighCVSS 8.1 EPSS 1.3%
Published
Updated
Exploit status
No public exploit known

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP server on the pump contains hardcoded credentials, which are not fully initialized. The FTP server is only accessible if the pump is configured to allow FTP connections.

Upgrade path

1 package: 0 fixed.

PackageAffectedFixed inStatus
smiths-medical medfusion 4000 wireless syringe infusion pump = 1.1; = 1.5; = 1.6 No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
smiths-medical medfusion 4000 wireless syringe infusion pump = 1.1 No fix listed
smiths-medical medfusion 4000 wireless syringe infusion pump = 1.5 No fix listed
smiths-medical medfusion 4000 wireless syringe infusion pump = 1.6 No fix listed

References