CVE-2017-12725 A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-c...
- Published
- Updated
- Exploit status
- No public exploit known
Description
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
Upgrade path
1 package: 0 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.1; = 1.5; = 1.6 | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.1 | No fix listed |
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.5 | No fix listed |
| smiths-medical medfusion 4000 wireless syringe infusion pump | = 1.6 | No fix listed |
References
- 100665 third-party-advisory vdb-entry
- ics-cert.us-cert.gov third-party-advisory us-government-resource