Skip to content

CVE-2017-5149 An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The...

HighCVSS 8.9 EPSS 1.2%
Published
Updated
Exploit status
No public exploit known

Description

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

Affected products

ProductAffected versionsFixed in
abbott merlinhome firmware <= 8.0 No fix listed

References