CVE-2018-18564 An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, and....
HighCVSS 7.4
EPSS 0.58%
- Published
- Updated
- Exploit status
- No public exploit known
Description
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). Improper access control allows attackers in the adjacent network to change the instrument configuration.
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| roche accu-chek inform ii firmware | < 03.06.00 | 03.06.00 |
| roche accu-chek inform ii firmware | >= 04.00.00, < 04.03.00 | 04.03.00 |
| roche cobas h 232 firmware | < 04.00.04 | 04.00.04 |
| roche coaguchek pro ii firmware | < 04.03.00 | 04.03.00 |
References
- 105843 third-party-advisory vdb-entry
- ics-cert.us-cert.gov mitigation third-party-advisory us-government-resource