Skip to content

CVE-2018-18565 An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, Coag...

MediumCVSS 6.8 EPSS 0.48%
Published
Updated
Exploit status
No public exploit known

Description

An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.

Affected products

ProductAffected versionsFixed in
roche accu-chek inform ii firmware < 03.06.00 03.06.00
roche accu-chek inform ii firmware >= 04.00.00, < 04.03.00 04.03.00
roche cobas h 232 firmware < 03.01.03 03.01.03
roche cobas h 232 firmware >= 04.00.00, < 04.00.04 04.00.04
roche coaguchek pro ii firmware < 04.03.00 04.03.00
roche coaguchek xs plus firmware < 03.01.06 03.01.06
roche coaguchek xs pro firmware < 03.01.06 03.01.06