CVE-2018-19010 Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. B...

MediumCVSS 6.5 EPSS 0.76%
Published
Updated
Exploit status
No public exploit known

Description

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.

Upgrade path

5 packages: 0 fixed, 4 with no fix yet.

PackageAffectedFixed inStatus
draeger delta xl firmware All versions No fix listed No fix yet
draeger infinity delta firmware All versions No fix listed No fix yet
draeger infinity explorer c700 firmware All versions No fix listed No fix yet
draeger kappa firmware All versions No fix listed No fix yet
ICS-CERT Dräger Infinity Delta = Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
draeger kappa firmware All versions No fix listed
draeger infinity explorer c700 firmware All versions No fix listed
draeger delta xl firmware All versions No fix listed
draeger infinity delta firmware All versions No fix listed
ICS-CERT Dräger Infinity Delta = Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. No fix listed