CVE-2018-4845 A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 S...
- Published
- Updated
- Exploit status
- No public exploit known
Description
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Upgrade path
4 packages: 1 fixed, 1 partially fixed, 1 with no fix yet.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Siemens rapidpoint 400 firmware | All versions | No fix listed | No fix yet |
| Siemens rapidpoint 500 firmware | <= 2.3; >= 3.0 | after 2.3 | Partially fixed |
| Siemens rapidlab 1200 firmware | < 3.3 | 3.3 | Fixed |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDLab 1200 Series : All versions < V3.3 _with_ Siemens Healthineers Informatics products; = RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products; = RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products; = RAPIDPoint 500 systems : All versions =< V2.3 _with_ Siemens Healthineers Informatics products +2 more | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Siemens rapidpoint 400 firmware | All versions | No fix listed |
| Siemens rapidpoint 500 firmware | <= 2.3 | No fix listed |
| Siemens rapidpoint 500 firmware | >= 3.0 | No fix listed |
| Siemens rapidlab 1200 firmware | < 3.3 | 3.3 |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems : All versions _without_ use of Siemens Healthineers Informatics products | No fix listed |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDLab 1200 Series : All versions < V3.3 _with_ Siemens Healthineers Informatics products | No fix listed |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDPoint 500 systems : All versions >= V3.0 _with_ Siemens Healthineers Informatics products | No fix listed |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDPoint 500 systems : V2.4.X _with_ Siemens Healthineers Informatics products | No fix listed |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDPoint 500 systems : All versions =< V2.3 _with_ Siemens Healthineers Informatics products | No fix listed |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDPoint 400 systems : All versions_with_ Siemens Healthineers Informatics products | No fix listed |
References
- cert-portal.siemens.com third-party-advisory