CVE-2018-4846 A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 S...
- Published
- Updated
- Exploit status
- No public exploit known
Description
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Siemens rapidpoint 400 firmware | All versions | No fix listed |
| Siemens rapidpoint 500 firmware | <= 2.3 | No fix listed |
| Siemens rapidpoint 500 firmware | >= 3.0 | No fix listed |
| Siemens rapidlab 1200 firmware | < 3.3 | 3.3 |
| Siemens RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems, RAPIDLab 1200 Series, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 500 systems, RAPIDPoint 400 systems | = RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions < V3.3 _with_Si | No fix listed |
References
- cert-portal.siemens.com third-party-advisory