CVE-2018-5438 Philips ISCV application prior to version 2.3.0 has an insufficient session expiration vulnerability where an attacker could reuse the session of a previously logged in user. This vulnerability exists...
MediumCVSS 6.3
EPSS 0.44%
- Published
- Updated
- Exploit status
- No public exploit known
Description
Philips ISCV application prior to version 2.3.0 has an insufficient session expiration vulnerability where an attacker could reuse the session of a previously logged in user. This vulnerability exists when using ISCV together with an Electronic Medical Record (EMR) system, where ISCV is in KIOSK mode for multiple users and using Windows authentication. This may allow an attacker to gain unauthorized access to patient health information and potentially modify this information.
Upgrade path
1 package: 1 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Philips intellispace cardiovascular | <= 2.3.0 | after 2.3.0 | Fixed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Philips intellispace cardiovascular | <= 2.3.0 | No fix listed |
References
- 102847 third-party-advisory vdb-entry
- ics-cert.us-cert.gov mitigation third-party-advisory us-government-resource
- www.usa.philips.com mitigation vendor-advisory