Skip to content

CVE-2018-8853 Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk....

HighCVSS 8.8 EPSS 0.36%
Published
Updated
Exploit status
No public exploit known
Technology
Windows

Description

Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.

Upgrade path

5 packages: 4 fixed.

PackageAffectedFixed inStatus
Philips brilliance ct big bore firmware <= 2.3.5 after 2.3.5 Fixed
Philips brilliance firmware 64 <= 2.6.2 after 2.6.2 Fixed
Philips brilliance ict firmware <= 4.1.6 after 4.1.6 Fixed
Philips brilliance ict sp firmware <= 3.2.4 after 3.2.4 Fixed
Philips Brilliance CT Scanners = Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
Philips brilliance firmware 64 <= 2.6.2 No fix listed
Philips brilliance ict sp firmware <= 3.2.4 No fix listed
Philips brilliance ict firmware <= 4.1.6 No fix listed
Philips brilliance ct big bore firmware <= 2.3.5 No fix listed
Philips Brilliance CT Scanners = Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. No fix listed

References