CVE-2018-8853 Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk....
HighCVSS 8.8
EPSS 0.36%
- Published
- Updated
- Exploit status
- No public exploit known
- Technology
- Windows
Description
Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.
Upgrade path
5 packages: 4 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Philips brilliance ct big bore firmware | <= 2.3.5 | after 2.3.5 | Fixed |
| Philips brilliance firmware 64 | <= 2.6.2 | after 2.6.2 | Fixed |
| Philips brilliance ict firmware | <= 4.1.6 | after 4.1.6 | Fixed |
| Philips brilliance ict sp firmware | <= 3.2.4 | after 3.2.4 | Fixed |
| Philips Brilliance CT Scanners | = Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Philips brilliance firmware 64 | <= 2.6.2 | No fix listed |
| Philips brilliance ict sp firmware | <= 3.2.4 | No fix listed |
| Philips brilliance ict firmware | <= 4.1.6 | No fix listed |
| Philips brilliance ct big bore firmware | <= 2.3.5 | No fix listed |
| Philips Brilliance CT Scanners | = Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. | No fix listed |
References
- 104088 third-party-advisory vdb-entry
- ics-cert.us-cert.gov third-party-advisory us-government-resource
- www.usa.philips.com vendor-advisory