CVE-2019-25719 Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling

HighCVSS 8.8 EPSS 0.13%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: yes · technical impact: partial

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

Upgrade path

2 packages: 2 fixed.

PackageAffectedFixed inStatus
Dräger Infinity Acute Care System <= VG4.0.3; <= VG4.1.1 after VG4.0.3, VG4.1.1 Fixed
Dräger Standalone Infinity M540 patient monitor <= VG4.1.1 after VG4.1.1 Fixed