CVE-2019-25723 Dräger Perseus A500 2.00-2.02 DoS via Medibus Interface
MediumCVSS 6.3
EPSS 0.24%
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: no · technical impact: partial
Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
Upgrade path
1 package: 0 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Dräger Perseus A500 | = 2.00; = 2.01; = 2.02 | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Dräger Perseus A500 | = 2.00 | No fix listed |
| Dräger Perseus A500 | = 2.01 | No fix listed |
| Dräger Perseus A500 | = 2.02 | No fix listed |
References
- static.draeger.com vendor-advisory
- www.vulncheck.com third-party-advisory