CVE-2020-12037 Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Pa...
HighCVSS 7.5
EPSS 0.50%
- Published
- Updated
- Exploit status
- No public exploit known
Description
Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device.
Upgrade path
2 packages: 1 fixed, 1 with no fix yet.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| baxter prismaflex firmware | All versions | No fix listed | No fix yet |
| baxter prismax firmware | < 3.0 | 3.0 | Fixed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| baxter prismaflex firmware | All versions | No fix listed |
| baxter prismax firmware | < 3.0 | 3.0 |
References
- www.us-cert.gov not-applicable
- us-cert.cisa.gov third-party-advisory