CVE-2020-15486 An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of t...

MediumCVSS 6.5 EPSS 0.53%
Published
Updated
Exploit status
No public exploit known

Description

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Upgrade path

1 package: 0 fixed.

PackageAffectedFixed inStatus
drtrust electrocardiogram pen firmware = 2.00.08 No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
drtrust electrocardiogram pen firmware = 2.00.08 No fix listed