Skip to content

CVE-2020-25187 Medtronic MyCareLink Smart Heap-based Buffer Overflow

CriticalCVSS 9.8 EPSS 3.9%
Published
Updated
Exploit status
No public exploit known

Description

Medtronic MyCareLink Smart 25000 is

 vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device

Upgrade path

2 packages: 0 fixed, 1 with no fix yet.

PackageAffectedFixed inStatus
medtronic mycarelink smart model 25000 firmware All versions No fix listed No fix yet
medtronic Smart Model 25000 Patient Reader = All versions No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
medtronic mycarelink smart model 25000 firmware All versions No fix listed
medtronic Smart Model 25000 Patient Reader = All versions No fix listed

References