CVE-2021-43935 ICSMA-21-343-01 Hillrom Welch Allyn Cardio Products
CriticalCVSS 9.8
EPSS 1.1%
- Published
- Updated
- Exploit status
- No public exploit known
Description
The impacted products, when configured to use SSO, are affected by an improper authentication vulnerability. This vulnerability allows the application to accept manual entry of any active directory (AD) account provisioned in the application without supplying a password, resulting in access to the application as the supplied AD account, with all associated privileges.
Upgrade path
14 packages: 12 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| baxter welch allyn connex cardio | >= 1.0.0, <= 1.1.1 | after 1.1.1 | Fixed |
| baxter welch allyn hscribe holter analysis system firmware | >= 5.01, <= 6.4.0 | after 6.4.0 | Fixed |
| baxter welch allyn q-stress cardiac stress testing system firmware | >= 6.0.0, <= 6.3.1 | after 6.3.1 | Fixed |
| baxter welch allyn rscribe resting ecg system | >= 5.01, <= 7.0.0 | after 7.0.0 | Fixed |
| baxter welch allyn vision express holter analysis system | >= 6.1.0, <= 6.4.0 | after 6.4.0 | Fixed |
| baxter welch allyn xscribe cardiac stress testing system firmware | >= 5.01, <= 6.3.1 | after 6.3.1 | Fixed |
| hillrom Welch Allyn Connex Cardio | >= 1.0.0, <= 1.1.1 | after 1.1.1 | Fixed |
| hillrom Welch Allyn H-Scribe Holter Analysis System | >= 5.01, <= 6.4.0 | after 6.4.0 | Fixed |
| hillrom Welch Allyn Q-Stress Cardiac Stress Testing System | >= 6.0.0, <= 6.3.1 | after 6.3.1 | Fixed |
| hillrom Welch Allyn R-Scribe Resting ECG System | >= 5.01, <= 7.0.0 | after 7.0.0 | Fixed |
| hillrom Welch Allyn Vision Express | >= 6.1.0, <= 6.4.0 | after 6.4.0 | Fixed |
| hillrom Welch Allyn X-Scribe Cardiac Stress Testing System | >= 5.01, <= 6.3.1 | after 6.3.1 | Fixed |
| baxter welch allyn diagnostic cardiology suite | = 2.1.0 | No fix listed | Versions listed |
| hillrom Welch Allyn Diagnostic Cardiology Suite | = 2.1.0 | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| baxter welch allyn connex cardio | >= 1.0.0, <= 1.1.1 | No fix listed |
| baxter welch allyn diagnostic cardiology suite | = 2.1.0 | No fix listed |
| baxter welch allyn rscribe resting ecg system | >= 5.01, <= 7.0.0 | No fix listed |
| baxter welch allyn vision express holter analysis system | >= 6.1.0, <= 6.4.0 | No fix listed |
| baxter welch allyn hscribe holter analysis system firmware | >= 5.01, <= 6.4.0 | No fix listed |
| baxter welch allyn q-stress cardiac stress testing system firmware | >= 6.0.0, <= 6.3.1 | No fix listed |
| baxter welch allyn xscribe cardiac stress testing system firmware | >= 5.01, <= 6.3.1 | No fix listed |
| hillrom Welch Allyn Q-Stress Cardiac Stress Testing System | >= 6.0.0, <= 6.3.1 | No fix listed |
| hillrom Welch Allyn X-Scribe Cardiac Stress Testing System | >= 5.01, <= 6.3.1 | No fix listed |
| hillrom Welch Allyn Diagnostic Cardiology Suite | = 2.1.0 | No fix listed |
| hillrom Welch Allyn Vision Express | >= 6.1.0, <= 6.4.0 | No fix listed |
| hillrom Welch Allyn H-Scribe Holter Analysis System | >= 5.01, <= 6.4.0 | No fix listed |
| hillrom Welch Allyn R-Scribe Resting ECG System | >= 5.01, <= 7.0.0 | No fix listed |
| hillrom Welch Allyn Connex Cardio | >= 1.0.0, <= 1.1.1 | No fix listed |
References
- www.cisa.gov mitigation third-party-advisory us-government-resource