CVE-2022-38069 Contec Health CMS8000
MediumCVSS 6.1
EPSS 0.34%
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: no · technical impact: partial
Description
Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters
Upgrade path
2 packages: 0 fixed, 2 with no fix yet.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All versions | No fix listed | No fix yet |
| contechealth cms8000 firmware | All versions | No fix listed | No fix yet |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Contec Health CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor | All versions | No fix listed |
| contechealth cms8000 firmware | All versions | No fix listed |
References
- www.cisa.gov mitigation third-party-advisory us-government-resource