CVE-2022-41627 The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could all...

HighCVSS 7.6 EPSS 0.12%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: no · technical impact: partial

Description

The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.

Upgrade path

4 packages: 0 fixed, 4 with no fix yet.

PackageAffectedFixed inStatus
AliveCor KardiaMobile All versions No fix listed No fix yet
AliveCor kardiamobile 6l firmware All versions No fix listed No fix yet
AliveCor kardiamobile card firmware All versions No fix listed No fix yet
AliveCor kardiamobile firmware All versions No fix listed No fix yet

Affected products

ProductAffected versionsFixed in
AliveCor KardiaMobile All versions No fix listed
AliveCor kardiamobile firmware All versions No fix listed
AliveCor kardiamobile 6l firmware All versions No fix listed
AliveCor kardiamobile card firmware All versions No fix listed