Description

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Affected products

ProductAffected versionsFixed in
B. Braun Space Battery Pack SP with Wi-Fi <= 053L000092 No fix listed
B. Braun Space Battery Pack SP with Wi-Fi <= 054U000092 No fix listed
bbraun battery-pack sp with wifi firmware <= 053l000092 No fix listed
bbraun battery-pack sp with wifi firmware <= 054u000092 No fix listed

References