CVE-2023-25931 Medtronic Micro Clinician & InterStim X Clinician App Password Reset Issue
MediumCVSS 6.8
EPSS 0.27%
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: no · technical impact: partial
Description
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy application, which has greater control over therapy parameters than the patient app. Changes still cannot be made outside of the established therapy parameters of the programmer. For unauthorized access to occur, an individual would need physical access to the Smart Programmer.
Upgrade path
3 packages: 1 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| medtronic InsterStim Applications | >= InterStim X Clinician, < A51300; >= Micro Clinician, < A51200 | A51200, A51300 | Fixed |
| medtronic interstim x clinician | = a51300 | No fix listed | Versions listed |
| medtronic micro clinician | = a51200 | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| medtronic InsterStim Applications | >= Micro Clinician, < A51200 | A51200 |
| medtronic InsterStim Applications | >= InterStim X Clinician, < A51300 | A51300 |
| medtronic interstim x clinician | = a51300 | No fix listed |
| medtronic micro clinician | = a51200 | No fix listed |
References
- global.medtronic.com vendor-advisory