Skip to content

CVE-2023-30561 Lack of Cryptographic Security of IUI Bus

MediumCVSS 6.1 EPSS 0.19%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: no · technical impact: partial

Description

The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.

Upgrade path

2 packages: 2 fixed.

PackageAffectedFixed inStatus
bd alaris 8015 pcu firmware <= 12.1.3 after 12.1.3 Fixed
Becton Dickinson & Co BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 <= 12.1.3 after 12.1.3 Fixed

Affected products

ProductAffected versionsFixed in
Becton Dickinson & Co BD Alarisâ„¢ Point-of-Care Unit (PCU) Model 8015 <= 12.1.3 No fix listed
bd alaris 8015 pcu firmware <= 12.1.3 No fix listed

References