Skip to content

CVE-2023-52952 A vulnerability has been identified in HiMed Cockpit 12 pro (J31032-K2017-H259) (All versions >= V11.5.1 < V11.6.2), HiMed Cockpit 14 pro+ (J31032-K2017-H435) (All versions >= V11.5.1 < V11.6.2), HiMe...

CriticalCVSS 9.3 EPSS 0.17%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: no · technical impact: partial

Description

A vulnerability has been identified in HiMed Cockpit 12 pro (J31032-K2017-H259) (All versions >= V11.5.1 < V11.6.2), HiMed Cockpit 14 pro+ (J31032-K2017-H435) (All versions >= V11.5.1 < V11.6.2), HiMed Cockpit 18 pro (J31032-K2017-H260) (All versions >= V11.5.1 < V11.6.2), HiMed Cockpit 18 pro+ (J31032-K2017-H436) (All versions >= V11.5.1 < V11.6.2). The Kiosk Mode of the affected devices contains a restricted desktop environment escape vulnerability. This could allow an unauthenticated local attacker to escape the restricted environment and gain access to the underlying operating system.

Upgrade path

4 packages: 4 fixed.

PackageAffectedFixed inStatus
Siemens HiMed Cockpit 12 pro >= V11.5.1, < V11.6.2 V11.6.2 Fixed
Siemens HiMed Cockpit 14 pro+ >= V11.5.1, < V11.6.2 V11.6.2 Fixed
Siemens HiMed Cockpit 18 pro >= V11.5.1, < V11.6.2 V11.6.2 Fixed
Siemens HiMed Cockpit 18 pro+ >= V11.5.1, < V11.6.2 V11.6.2 Fixed

Affected products

ProductAffected versionsFixed in
Siemens HiMed Cockpit 12 pro >= V11.5.1, < V11.6.2 V11.6.2
Siemens HiMed Cockpit 14 pro+ >= V11.5.1, < V11.6.2 V11.6.2
Siemens HiMed Cockpit 18 pro >= V11.5.1, < V11.6.2 V11.6.2
Siemens HiMed Cockpit 18 pro+ >= V11.5.1, < V11.6.2 V11.6.2

References