CVE-2024-10476 Default credentials are used in the above listed BD Diagnostic Solutions products. If exploited, threat actors may be able to access, modify or delete data, including sensitive information such as pro...
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: no · technical impact: total
Description
Default credentials are used in the above listed BD Diagnostic Solutions products. If exploited, threat actors may be able to access, modify or delete data, including sensitive information such as protected health information (PHI) and personally identifiable information (PII). Exploitation of this vulnerability may allow an attacker to shut down or otherwise impact the availability of the system. Note: BD Synapsys™ Informatics
Solution is only in scope of
this vulnerability when
installed on a NUC server. BD Synapsys™
Informatics Solution installed
on a customer-provided virtual machine or on the BD Kiestra™ SCU hardware is
not in scope.
Upgrade path
6 packages: 6 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Becton Dickinson & Co BD BACTEC™ Blood Culture System | <= 7.20 | after 7.20 | Fixed |
| Becton Dickinson & Co BD COR™ System | <= 8.90 | after 8.90 | Fixed |
| Becton Dickinson & Co BD EpiCenter™ Microbiology Data Management System | <= 7.45 | after 7.45 | Fixed |
| Becton Dickinson & Co BD MAX™ System | <= 6.10 | after 6.10 | Fixed |
| Becton Dickinson & Co BD Phoenix™ M50 Automated Microbiology System | <= 2.70 | after 2.70 | Fixed |
| Becton Dickinson & Co BD Synapsys™ Informatics Solution | <= 6.10 | after 6.10 | Fixed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Becton Dickinson & Co BD BACTEC™ Blood Culture System | <= 7.20 | No fix listed |
| Becton Dickinson & Co BD COR™ System | <= 8.90 | No fix listed |
| Becton Dickinson & Co BD EpiCenter™ Microbiology Data Management System | <= 7.45 | No fix listed |
| Becton Dickinson & Co BD MAX™ System | <= 6.10 | No fix listed |
| Becton Dickinson & Co BD Phoenix™ M50 Automated Microbiology System | <= 2.70 | No fix listed |
| Becton Dickinson & Co BD Synapsys™ Informatics Solution | <= 6.10 | No fix listed |