CVE-2024-12248 Out-of-bounds Write vulnerability in Contec Health CMS8000 Patient Monitor
CriticalCVSS 9.3
EPSS 1.3%
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: yes · technical impact: total
Description
Contec Health CMS8000 Patient Monitor is vulnerable to an out-of-bounds write, which could allow an attacker to send specially formatted UDP requests in order to write arbitrary data. This could result in remote code execution.
Upgrade path
1 package: 0 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Contec Health CMS8000 Patient Monitor | = Firmware version CMS7.820.075.08/0.74(0.75); = Firmware version CMS7.820.120.01/0.93(0.95); = Firmware version smart3250-2.6.27-wlan2.1.7.cramfs | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Contec Health CMS8000 Patient Monitor | = Firmware version smart3250-2.6.27-wlan2.1.7.cramfs | No fix listed |
| Contec Health CMS8000 Patient Monitor | = Firmware version CMS7.820.075.08/0.74(0.75) | No fix listed |
| Contec Health CMS8000 Patient Monitor | = Firmware version CMS7.820.120.01/0.93(0.95) | No fix listed |