CVE-2024-48966 Life2000 service tools for test and calibration do not support user authentication

CriticalCVSS 10.0 EPSS 0.70%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: yes · technical impact: total

Description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

Upgrade path

2 packages: 1 fixed.

PackageAffectedFixed inStatus
baxter life2000 ventilator firmware < 6.08.00.00 6.08.00.00 Fixed
baxter Life2000 Ventilation System = 06.08.00.00 and prior No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
baxter Life2000 Ventilation System = 06.08.00.00 and prior No fix listed
baxter life2000 ventilator firmware < 6.08.00.00 6.08.00.00