CVE-2024-48971 Clinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventilator
CriticalCVSS 9.3
EPSS 0.22%
- Published
- Updated
- Exploit status
- No public exploit known
Description
The Clinician Password and Serial Number Clinician Password are hard-coded into the ventilator in plaintext form. This could allow an attacker to obtain the password off the ventilator and use it to gain unauthorized access to the device, with clinician privileges.
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| baxter Life2000 Ventilation System | = 06.08.00.00 and prior | No fix listed |
| baxter life2000 ventilator firmware | <= 06.08.00.00 | No fix listed |