CVE-2024-9832 No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator
CriticalCVSS 9.3
EPSS 0.18%
- Published
- Updated
- Exploit status
- No public exploit known
Description
There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| baxter Life2000 Ventilation System | = 06.08.00.00 and prior | No fix listed |
| baxter life2000 ventilator firmware | <= 06.08.00.00 | No fix listed |