CVE-2025-14096 Credential Disclosure vulnerability in Radiometer Products

HighCVSS 8.4 EPSS 0.22%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: no · technical impact: total

Description

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.

Other related CVE's are CVE-2025-14095 & CVE-2025-14097.

Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure:

Attacker requires physical access to the analyzer.

Temporary work Around:
Only authorized people can physically access the analyzer.

Permanent solution:
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.

Exploit Status:

Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Upgrade path

3 packages: 0 fixed.

PackageAffectedFixed inStatus
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers = Windows 7 Operating system; = Windows XP Operating system No fix listed Versions listed
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers = Windows 7 Operating system; = Windows XP Operating system No fix listed Versions listed
Radiometer Medical Aps AQT90 FLEX Analyzers = Windows 7 Operating system; = Windows XP Operating system No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers = Windows 7 Operating system No fix listed
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers = Windows XP Operating system No fix listed
Radiometer Medical Aps AQT90 FLEX Analyzers = Windows 7 Operating system No fix listed
Radiometer Medical Aps AQT90 FLEX Analyzers = Windows XP Operating system No fix listed
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers = Windows 7 Operating system No fix listed
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers = Windows XP Operating system No fix listed