CVE-2025-14097 Remote Code Execution Vulnerability in Radiometer Products

HighCVSS 7.2 EPSS 0.93%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: no · technical impact: total

Description

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.

Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution:
Customers should ensure the following:
• The network is secure, and access follows best practices.
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                     
Exploit Status:

Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Upgrade path

3 packages: 0 fixed.

PackageAffectedFixed inStatus
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers = Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems No fix listed Versions listed
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers = Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems No fix listed Versions listed
Radiometer Medical Aps AQT90 FLEX Analyzers = Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems No fix listed Versions listed

Affected products

ProductAffected versionsFixed in
Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers = Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems No fix listed
Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers = Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems No fix listed
Radiometer Medical Aps AQT90 FLEX Analyzers = Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems No fix listed

References