CVE-2025-14097 Remote Code Execution Vulnerability in Radiometer Products
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: no · technical impact: total
Description
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.
Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution:
Customers should ensure the following:
• The network is secure, and access follows best practices.
Local Radiometer representatives will contact all affected customers to discuss a permanent solution.
Exploit Status:
Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Upgrade path
3 packages: 0 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers | = Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems | No fix listed | Versions listed |
| Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers | = Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems | No fix listed | Versions listed |
| Radiometer Medical Aps AQT90 FLEX Analyzers | = Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems | No fix listed | Versions listed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers | = Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems | No fix listed |
| Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers | = Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems | No fix listed |
| Radiometer Medical Aps AQT90 FLEX Analyzers | = Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems | No fix listed |