CVE-2025-15653 Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
HighCVSS 7.0
EPSS 0.17%
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: no · technical impact: total
Description
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
Upgrade path
2 packages: 2 fixed.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| Dräger Zeus IE | < 1.0.5 | 1.0.5 | Fixed |
| Dräger Zeus RS C500 | < 1.0.9 | 1.0.9 | Fixed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Dräger Zeus IE | < 1.0.5 | 1.0.5 |
| Dräger Zeus RS C500 | < 1.0.9 | 1.0.9 |
References
- static.draeger.com vendor-advisory
- www.vulncheck.com third-party-advisory