Skip to content

CVE-2026-17583 Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check

HighCVSS 8.3 EPSS 0.14%
Published
Updated
Exploit status
No public exploit known
CISA SSVC
No known exploitation · automatable: no · technical impact: total
Technology
Ubuntu

Description

The affected

Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.

Upgrade path

20 packages: 16 fixed, 4 with no fix yet.

PackageAffectedFixed inStatus
ubuntu/genetic(Ubuntu 22.04) All versions No fix listed No fix yet
ubuntu/genetic(Ubuntu 24.04) All versions No fix listed No fix yet
ubuntu/genetic(Ubuntu 26.04) All versions No fix listed No fix yet
ubuntu/genetic(Ubuntu 26.10) All versions No fix listed No fix yet
Thermo Fisher ABI PRISM 310 Data Collection Software <= 3.1 after 3.1 Fixed
Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software <= 2.0 after 2.0 Fixed
Thermo Fisher Applied Biosystems 3130 Series Data Collection Software <= 4.1 after 4.1 Fixed
Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software <= 4.0.2 after 4.0.2 Fixed
Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software <= 5.0.2 after 5.0.2 Fixed
Thermo Fisher Applied Biosystems GeneMapper ID-X Software <= 1.7.3 after 1.7.3 Fixed
Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software <= 1.2.0 after 1.2.0 Fixed
Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software <= 1.2.5 after 1.2.5 Fixed
thermofisher abi prism 310 data collection software <= 3.1 after 3.1 Fixed
thermofisher abi prism 31003100-avant data collection software <= 2.0 after 2.0 Fixed
thermofisher applied biosystems 3130 series data collection software <= 4.1 after 4.1 Fixed
thermofisher applied biosystems 35003500xl series data collection software < 4.0.3 4.0.3 Fixed
thermofisher applied biosystems 37303730xl series data collection software < 5.0.3 5.0.3 Fixed
thermofisher applied biosystems seqstudio flex series instrument software < 1.2.1 1.2.1 Fixed
thermofisher applied biosystems seqstudio genetic analyzer data collection software < 1.2.6 1.2.6 Fixed
thermofisher genemapper id-x < 1.7.4 1.7.4 Fixed

Affected products

ProductAffected versionsFixed in
Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software <= 4.0.2 No fix listed
Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software <= 5.0.2 No fix listed
Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software <= 1.2.5 No fix listed
Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software <= 1.2.0 No fix listed
Thermo Fisher Applied Biosystems GeneMapper ID-X Software <= 1.7.3 No fix listed
Thermo Fisher Applied Biosystems 3130 Series Data Collection Software <= 4.1 No fix listed
Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software <= 2.0 No fix listed
Thermo Fisher ABI PRISM 310 Data Collection Software <= 3.1 No fix listed
ubuntu/genetic(Ubuntu:22.04) All versions No fix listed
ubuntu/genetic(Ubuntu:24.04) All versions No fix listed
ubuntu/genetic(Ubuntu:26.04) All versions No fix listed
ubuntu/genetic(Ubuntu:26.10) All versions No fix listed
thermofisher abi prism 310 data collection software <= 3.1 No fix listed
thermofisher abi prism 31003100-avant data collection software <= 2.0 No fix listed
thermofisher applied biosystems 3130 series data collection software <= 4.1 No fix listed
thermofisher applied biosystems 35003500xl series data collection software < 4.0.3 4.0.3
thermofisher applied biosystems 37303730xl series data collection software < 5.0.3 5.0.3
thermofisher applied biosystems seqstudio flex series instrument software < 1.2.1 1.2.1
thermofisher applied biosystems seqstudio genetic analyzer data collection software < 1.2.6 1.2.6
thermofisher genemapper id-x < 1.7.4 1.7.4

References