CVE-2026-17583 Thermo Fisher Applied Biosystems Genetic Analyzers Missing Support for Integrity Check
HighCVSS 8.3
EPSS 0.14%
- Published
- Updated
- Exploit status
- No public exploit known
- CISA SSVC
- No known exploitation · automatable: no · technical impact: total
- Technology
- Ubuntu
Description
The affected
Thermo Fisher Applied Biosystems Genetic Analyzers are vulnerable because .fsa/.hid output files can be edited. An attacker could tamper with these files, altering DNA data and resulting in inaccurate DNA test outcomes.
Upgrade path
20 packages: 16 fixed, 4 with no fix yet.
| Package | Affected | Fixed in | Status |
|---|---|---|---|
| ubuntu/genetic(Ubuntu 22.04) | All versions | No fix listed | No fix yet |
| ubuntu/genetic(Ubuntu 24.04) | All versions | No fix listed | No fix yet |
| ubuntu/genetic(Ubuntu 26.04) | All versions | No fix listed | No fix yet |
| ubuntu/genetic(Ubuntu 26.10) | All versions | No fix listed | No fix yet |
| Thermo Fisher ABI PRISM 310 Data Collection Software | <= 3.1 | after 3.1 | Fixed |
| Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software | <= 2.0 | after 2.0 | Fixed |
| Thermo Fisher Applied Biosystems 3130 Series Data Collection Software | <= 4.1 | after 4.1 | Fixed |
| Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software | <= 4.0.2 | after 4.0.2 | Fixed |
| Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software | <= 5.0.2 | after 5.0.2 | Fixed |
| Thermo Fisher Applied Biosystems GeneMapper ID-X Software | <= 1.7.3 | after 1.7.3 | Fixed |
| Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software | <= 1.2.0 | after 1.2.0 | Fixed |
| Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software | <= 1.2.5 | after 1.2.5 | Fixed |
| thermofisher abi prism 310 data collection software | <= 3.1 | after 3.1 | Fixed |
| thermofisher abi prism 31003100-avant data collection software | <= 2.0 | after 2.0 | Fixed |
| thermofisher applied biosystems 3130 series data collection software | <= 4.1 | after 4.1 | Fixed |
| thermofisher applied biosystems 35003500xl series data collection software | < 4.0.3 | 4.0.3 | Fixed |
| thermofisher applied biosystems 37303730xl series data collection software | < 5.0.3 | 5.0.3 | Fixed |
| thermofisher applied biosystems seqstudio flex series instrument software | < 1.2.1 | 1.2.1 | Fixed |
| thermofisher applied biosystems seqstudio genetic analyzer data collection software | < 1.2.6 | 1.2.6 | Fixed |
| thermofisher genemapper id-x | < 1.7.4 | 1.7.4 | Fixed |
Affected products
| Product | Affected versions | Fixed in |
|---|---|---|
| Thermo Fisher Applied Biosystems 3500/3500xL Series Data Collection Software | <= 4.0.2 | No fix listed |
| Thermo Fisher Applied Biosystems 3730/3730xL Series Data Collection Software | <= 5.0.2 | No fix listed |
| Thermo Fisher Applied Biosystems SeqStudio Genetic Analyzer Data Collection Software | <= 1.2.5 | No fix listed |
| Thermo Fisher Applied Biosystems SeqStudio Flex Series Instrument Software | <= 1.2.0 | No fix listed |
| Thermo Fisher Applied Biosystems GeneMapper ID-X Software | <= 1.7.3 | No fix listed |
| Thermo Fisher Applied Biosystems 3130 Series Data Collection Software | <= 4.1 | No fix listed |
| Thermo Fisher ABI PRISM 3100/3100-Avant Data Collection Software | <= 2.0 | No fix listed |
| Thermo Fisher ABI PRISM 310 Data Collection Software | <= 3.1 | No fix listed |
| ubuntu/genetic(Ubuntu:22.04) | All versions | No fix listed |
| ubuntu/genetic(Ubuntu:24.04) | All versions | No fix listed |
| ubuntu/genetic(Ubuntu:26.04) | All versions | No fix listed |
| ubuntu/genetic(Ubuntu:26.10) | All versions | No fix listed |
| thermofisher abi prism 310 data collection software | <= 3.1 | No fix listed |
| thermofisher abi prism 31003100-avant data collection software | <= 2.0 | No fix listed |
| thermofisher applied biosystems 3130 series data collection software | <= 4.1 | No fix listed |
| thermofisher applied biosystems 35003500xl series data collection software | < 4.0.3 | 4.0.3 |
| thermofisher applied biosystems 37303730xl series data collection software | < 5.0.3 | 5.0.3 |
| thermofisher applied biosystems seqstudio flex series instrument software | < 1.2.1 | 1.2.1 |
| thermofisher applied biosystems seqstudio genetic analyzer data collection software | < 1.2.6 | 1.2.6 |
| thermofisher genemapper id-x | < 1.7.4 | 1.7.4 |
References
- documents.thermofisher.com patch vendor-advisory
- www.cisa.gov patch third-party-advisory us-government-resource
- www.cve.org third-party-advisory
- Ubuntu CVE tracker vendor-advisory
- ENISA EUVD EUVD-2026-53616 third-party-advisory