An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.
Vendor
bostonscientific vulnerabilities
7 published CVES affecting bostonscientific, from public sources (CVE.org, NVD, CISA KEV, FIRST EPSS, OSV).
- Total
- 7
- Critical
- 0
- High
- 1
- Medium
- 6
- Low
- 0
- Unscored
- 0
- CISA KEV
- 0
Products
- zoom latitude pogrammerrecordermonitor 3120 firmware 5
- zoom latitude prm 3120 firmware 2
- zoom latitude programming system model 3120 firmware 1
Latest bostonscientific CVEs
Newest publication first
-
MediumCVSS 6.8 EPSS 0.15%
-
MediumCVSS 6.8 EPSS 0.17%
The programmer installation utility does not perform a cryptographic authenticity or integrity checks of the software on the flash drive. An attacker could leverage this weakness to install unauthorized software using a specially crafted USB.
-
HighCVSS 7.6 EPSS 0.26%
A skilled attacker with physical access to the affected device can gain access to the hard disk drive of the device to change the telemetry region and could use this setting to interrogate or program an implantable device in any region in the world.
-
CVE-2021-38398 Reliance on Component that is not Updateable for Boston Scientific Zoom Latitude
PublishedMediumCVSS 6.8 EPSS 0.43%The affected device uses off-the-shelf software components that contain unpatched vulnerabilities. A malicious attacker with physical access to the affected device could exploit these vulnerabilities.
-
MediumCVSS 6.4 EPSS 0.23%
An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted.
-
MediumCVSS 4.6 EPSS 0.39%
Boston Scientific ZOOM LATITUDE PRM Model 3120 uses a hard-coded cryptographic key to encrypt PHI prior to having it transferred to removable media. CVSS v3 base score: 4.6; CVSS vector string: AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
-
MediumCVSS 4.6 EPSS 0.27%
Boston Scientific ZOOM LATITUDE PRM Model 3120 does not encrypt PHI at rest. CVSS v3 base score: 4.6; CVSS vector string: AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N.
Sources: CVE.org, NVD, CISA KEV, FIRST EPSS, OSV and public exploit-reference metadata. Product matching comes from the published CVE records. A proof-of-concept means public code exists; it is not evidence of exploitation.