Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information witho...
Vendor
Abbott Laboratories vulnerabilities
3 published CVES affecting Abbott Laboratories, from public sources (CVE.org, NVD, CISA KEV, FIRST EPSS, OSV).
- Total
- 3
- Critical
- 0
- High
- 1
- Medium
- 2
- Low
- 0
- Unscored
- 0
- CISA KEV
- 0
Products
Latest Abbott Laboratories CVEs
Newest publication first
-
MediumCVSS 6.5 EPSS 0.24%
-
MediumCVSS 6.5 EPSS 0.63%
Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base sc...
-
HighCVSS 8.8 EPSS 1.0%
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communi...