An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the...
Product
bbraun spacecom2 vulnerabilities
5 published CVES affecting bbraun spacecom2, from public sources (CVE.org, NVD, CISA KEV, FIRST EPSS, OSV).
- Total
- 5
- Critical
- 2
- High
- 3
- Medium
- 0
- Low
- 0
- Unscored
- 0
- CISA KEV
- 0
Latest bbraun spacecom2 CVEs
Newest publication first
-
HighCVSS 8.8 EPSS 0.83%
-
CriticalCVSS 9.8 EPSS 5.6%
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and exec...
-
CriticalCVSS 9.1 EPSS 0.98%
An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows remote attackers to upload any files to the /tmp directory of the device through the webpage API. This can result in critical files being overwritten.
-
HighCVSS 7.5 EPSS 0.83%
A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration.
-
HighCVSS 8.6 EPSS 1.1%
A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.
Sources: CVE.org, NVD, CISA KEV, FIRST EPSS, OSV and public exploit-reference metadata. Product matching comes from the published CVE records. A proof-of-concept means public code exists; it is not evidence of exploitation.