Skip to content

Vendor

bbraun vulnerabilities

21 published CVES affecting bbraun, from public sources (CVE.org, NVD, CISA KEV, FIRST EPSS, OSV).

Total
21
Critical
3
High
12
Medium
5
Low
1
Unscored
0
CISA KEV
0
Atom feed of bbraun alerts

Products

Latest bbraun CVEs

Newest publication first

  1. HighCVSS 8.8 EPSS 0.83%

    An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the...

  2. CriticalCVSS 9.8 EPSS 5.6%

    An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and exec...

  3. CriticalCVSS 9.1 EPSS 0.98%

    An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows remote attackers to upload any files to the /tmp directory of the device through the webpage API. This can result in critical files being overwritten.

  4. HighCVSS 7.5 EPSS 0.83%

    A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration.

  5. HighCVSS 8.6 EPSS 1.1%

    A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.

  6. CriticalCVSS 9.8 EPSS 2.0%

    A relative path traversal attack in the B. Braun OnlineSuite Version AP 3.0 and earlier allows unauthenticated attackers to upload or download arbitrary files.

  7. HighCVSS 7.8 EPSS 0.47%

    A DLL hijacking vulnerability in the B. Braun OnlineSuite Version AP 3.0 and earlier allows local attackers to execute code on the system as a high privileged user.

  8. HighCVSS 7.8 EPSS 0.98%

    An Excel Macro Injection vulnerability exists in the export feature in the B. Braun OnlineSuite Version AP 3.0 and earlier via multiple input fields that are mishandled in an Excel export.

  9. MediumCVSS 6.1 EPSS 0.95%

    An open redirect issue was discovered in B. Braun Medical SpaceCom module, which is integrated into the SpaceStation docking station: SpaceStation with SpaceCom module (integrated as part number 8713142U), software versions prior to Version 012U000040, and SpaceStation (part numb...