Medtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials to modify encrypted drive data.
Vendor
medtronic vulnerabilities
35 published CVES affecting medtronic, from public sources (CVE.org, NVD, CISA KEV, FIRST EPSS, OSV).
- Total
- 35
- Critical
- 2
- High
- 7
- Medium
- 25
- Low
- 1
- Unscored
- 0
- CISA KEV
- 0
Products
- MyCareLink Patient Monitor 24950 5
- MyCareLink Patient Monitor 24952 5
- 24950 MyCareLink Monitor 4
- 24952 MyCareLink Monitor 4
- 29901 Encore Programmer 4
- CareLink Network 4
- Valleylab FT10 Energy Platform (VLFT10GEN) 4
- valleylab ft10 energy platform firmware 4
- 2090 CareLink Programmer 3
- 2090 carelink programmer firmware 3
- CareLink 2090 Programmer 3
- mycarelink smart model 25000 firmware 3
- Smart Model 25000 Patient Reader 3
- 24950 mycarelink monitor firmware 2
- 24952 mycarelink monitor firmware 2
- 8870 N’Vision removable Application Card 2
- Amplia CRT-D 2
- amplia crt-d firmware 2
- Brava CRT-D 2
- carelink 2090 firmware 2
- CareLink Monitor 2
- Claria CRT-D 2
- claria crt-d firmware 2
- Compia CRT-D 2
- compia crt-d firmware 2
- Concerto CRT-D 2
- concerto crt-d firmware 2
- Concerto II CRT-D 2
- concerto ii crt-d firmware 2
- Conexus Radio Frequency Telemetry Protocol 2
Latest medtronic CVEs
Newest publication first · page 1 of 2
-
MediumCVSS 6.8 EPSS 0.13%
-
MediumCVSS 6.8 EPSS 0.16%
Medtronic MyCareLink Patient Monitor has an internal serial interface, which allows an attacker with physical access to access a login prompt via a UART terminal.
-
LowCVSS 3.1 EPSS 0.20%
Insecure Direct Object Reference vulnerability in Medtronic CareLink Network which allows an authenticated attacker with access to specific device and user information to submit web requests to an API endpoint that would expose sensitive user information. This issue affects CareL...
-
MediumCVSS 4.1 EPSS 0.12%
Medtronic CareLink Network allows a local attacker with access to log files on an internal API server to view plaintext passwords from errors logged under certain circumstances. This issue affects CareLink Network: before December 4, 2025.
-
CriticalCVSS 9.8 EPSS 0.36%
Medtronic CareLink Network allows an unauthenticated remote attacker to perform a brute force attack on an API endpoint that could be used to determine a valid password under certain circumstances. This issue affects CareLink Network: before December 4, 2025.
-
MediumCVSS 5.3 EPSS 0.32%
Medtronic CareLink Network allows an unauthenticated remote attacker to initiate a request for security questions to an API endpoint that could be used to determine a valid user account. This issue affects CareLink Network: before December 4, 2025.
-
MediumCVSS 6.8 EPSS 0.31%
Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before J...
-
MediumCVSS 6.8 EPSS 0.22%
Medtronic MyCareLink Patient Monitor uses an unencrypted filesystem on internal storage, which allows an attacker with physical access to read and modify files. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
-
MediumCVSS 6.5 EPSS 0.19%
Medtronic MyCareLink Patient Monitor has an internal service that deserializes data, which allows a local attacker to interact with the service by crafting a binary payload to crash the service or elevate privileges. This issue affects MyCareLink Patient Monitor models 24950 and...
-
HighCVSS 8.8 EPSS 28.3%
Deserialization of untrusted data in Microsoft Messaging Queuing Service in Medtronic's Paceart Optima versions 1.11 and earlier on Windows allows an unauthorized user to impact a healthcare delivery organization’s Paceart Optima system cardiac device causing data to be deleted,...
-
MediumCVSS 6.8 EPSS 0.27%
Medtronic identified that the Pelvic Health clinician apps, which are installed on the Smart Programmer mobile device, have a password vulnerability that requires a security update to fix. Not updating could potentially result in unauthorized control of the clinician therapy appl...
-
MediumCVSS 4.8 EPSS 0.36%
A vulnerability exists which could allow an unauthorized user to learn aspects of the communication protocol used to pair system components while the pump is being paired with other system components. Exploitation requires nearby wireless signal proximity with the patient and the...
-
HighCVSS 8.1 EPSS 3.8%
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Pati...
-
CriticalCVSS 9.8 EPSS 3.9%
Medtronic MyCareLink Smart 25000 is vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely ex...
-
HighCVSS 8.8 EPSS 0.80%
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use anoth...
-
MediumCVSS 4.6 EPSS 0.39%
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the...
-
MediumCVSS 4.6 EPSS 0.37%
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, all...
-
HighCVSS 7.8 EPSS 0.27%
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interact...
-
HighCVSS 7.5 EPSS 2.0%
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use multiple sets of hard-coded credentials. If discovered, they c...
-
CVE-2019-10964 Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control
PublishedHighCVSS 7.1 EPSS 1.2%Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authoriza...
-
MediumCVSS 6.5 EPSS 0.19%
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D...
-
CVE-2019-6538 Medtronic Conexus Radio Frequency Telemetry Protocol Improper Access Control
PublishedMediumCVSS 6.5 EPSS 0.90%The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D...
-
MediumCVSS 4.6 EPSS 0.32%
Medtronic CareLink and Encore Programmers do not encrypt or do not sufficiently encrypt sensitive PII and PHI information while at rest .
-
MediumCVSS 5.3 EPSS 0.71%
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pum...
-
MediumCVSS 4.8 EPSS 0.47%
Communications between Medtronic MiniMed MMT pumps and wireless accessories are transmitted in cleartext. A sufficiently skilled attacker could capture these transmissions and extract sensitive information, such as device serial numbers.
Sources: CVE.org, NVD, CISA KEV, FIRST EPSS, OSV and public exploit-reference metadata. Product matching comes from the published CVE records. A proof-of-concept means public code exists; it is not evidence of exploitation.