Hospira MedNet software version 5.8 and prior uses vulnerable versions of the JBoss Enterprise Application Platform software that may allow unauthenticated users to execute arbitrary code on the target system. Hospira has developed a new version of the MedNet software, MedNet 6.1...
Vendor
hospira vulnerabilities
17 published CVES affecting hospira, from public sources (CVE.org, NVD, CISA KEV, FIRST EPSS, OSV).
- Total
- 17
- Critical
- 4
- High
- 9
- Medium
- 3
- Low
- 1
- Unscored
- 0
- CISA KEV
- 0
Products
- lifecare pcainfusion firmware 6
- Symbiq Infusion System 5
- mednet 4
- Plum A+ Infusion System 4
- Plum A+3 Infusion System 4
- lifecare pca infusion system 3
- lifecare pca3 2
- lifecare pca5 2
- communication engine 1
Latest hospira CVEs
Newest publication first
-
CriticalCVSS 9.8 EPSS 5.0%
-
HighCVSS 7.5 EPSS 0.77%
Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used...
-
CriticalCVSS 9.8 EPSS 0.95%
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices...
-
CriticalCVSS 9.8 EPSS 2.0%
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pu...
-
CriticalCVSS 9.8 EPSS 2.0%
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the aff...
-
HighCVSS 7.5 EPSS 1.0%
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the aff...
-
HighCVSS 8.8 EPSS 2.9%
Hospira Symbiq Infusion System 3.13 and earlier allows remote authenticated users to trigger "unanticipated operations" by leveraging "elevated privileges" for an unspecified call to an incorrectly exposed function.
-
HighCVSS 7.3 EPSS 1.1%
Stack-based buffer overflow in Hospira Communication Engine (CE) before 1.2 in LifeCare PCA Infusion System 5.07, Plum A+ Infusion System 13.40, and Plum A+3 Infusion System 13.40 allows remote attackers to cause a denial of service or possibly have unspecified other impact via t...
-
HighCVSS 7.8 EPSS 2.6%
Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to cause a denial of service (forced manual reboot) via a flood of TCP packets.
-
MediumCVSS 4.6 EPSS 0.54%
Hospira LifeCare PCA Infusion System before 7.0 stores private keys and certificates, which has unspecified impact and attack vectors.
-
HighCVSS 10.0 EPSS 6.6%
Stack-based buffer overflow in Hospira LifeCare PCA Infusion System 5.0 and earlier, and possibly other versions, allows remote attackers to execute arbitrary code via unspecified vectors.
-
MediumCVSS 5.0 EPSS 2.2%
Hospira LifeCare PCA Infusion System before 7.0 has hardcoded credentials, which makes it easier for remote attackers to obtain access via unspecified vectors.
-
HighCVSS 9.3 EPSS 1.2%
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (...
-
HighCVSS 10.0 EPSS 5.1%
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
-
HighCVSS 9.0 EPSS 2.3%
Hospira MedNet before 6.1 uses a hardcoded cleartext password to control SQL database authorization, which allows remote authenticated users to bypass intended access restrictions by leveraging knowledge of this password.
-
MediumCVSS 5.0 EPSS 1.7%
Hospira MedNet before 6.1 uses hardcoded cryptographic keys for protection of data transmission from infusion pumps, which allows remote attackers to obtain sensitive information by sniffing the network.
-
LowCVSS 2.1 EPSS 0.33%
The installation component in Hospira MedNet before 6.1 places cleartext credentials in configuration files, which allows local users to obtain sensitive information by reading a file.
Sources: CVE.org, NVD, CISA KEV, FIRST EPSS, OSV and public exploit-reference metadata. Product matching comes from the published CVE records. A proof-of-concept means public code exists; it is not evidence of exploitation.